MRI
MRI India Journals Vol. 15 No. 1 (2026)

Lower Urinary Tract Symptoms Diagnosis by Uropine: A MEMS based economic urodynamic system

Authors

  • Parth Singh SSIPMT, Raipur, Chhattisgarh.
  • Prince Raj SSIPMT, Raipur, Chhattisgarh.
  • Animesh Tiwari SSIPMT, Raipur, Chhattisgarh.
  • Golu Kumar Pal SSIPMT, Raipur, Chhattisgarh.
  • Vikas Singh Thakur SSIPMT, Raipur, Chhattisgarh.
  • Lekha Sahu SSIPMT, Raipur, Chhattisgarh.
  • Keshika Jangde SSIPMT, Raipur, Chhattisgarh.
  • Tegendra Sahu Department of Computer Science and Engineering, SSIPMT, Raipur, Chhattisgarh.
  • Indrajeet Bhalla Department of Electronics and Tele-communications, SSIPMT, Raipur, Chhattisgarh.
  • J. Gnanaraj Founder of RSI, Director RSI, SSIPMT, Raipur, Chhattisgarh.
  • Om Prakash Sahu Professor and Dean Integrated Research Centre, SSIPMT, Raipur, Chhattisgarh.

DOI:

https://doi.org/10.65521/intjournalrecadvengtech.v15i1.2065

Keywords:

Lower urinary tract symptoms (LUTS) Urodynamic study (UDS) Portable biomedical device MEMS pressure sensors AI-based diagnostics

Abstract

Lower urinary tract symptoms, or LUTS, are a major and common urological burden that impacts a wide range of people of all ages and genders. Two potential causes of LUTS are neurogenic bladder dysfunction and benign prostatic hyperplasia (BPH) [1].  Urodynamic study (UDS), a functional evaluation of bladder and urethra, is a standard diagnostic procedure for these conditions[2]. However, high cost, lack of portability, and operational complexity of multi-lumen catheterization and laborious fluid-filled pressure transmission lines are characteristics of conventional urodynamic instrumentation[2][3]. These restrictions significantly limit urological care accessibility, especially in rural and resource-constrained settings. Thorough hardware design, engineering process, and clinical validation of "Uropine," a novel, inexpensive, AI-powered portable biomedical device intended to upend a established paradigm of LUTS diagnosis, are described in this research paper.

Uropine system provides real-time, high-fidelity monitoring of intra-vesical pressure (IBP) and intra-abdominal pressure (IAP) by combining sophisticated Micro-Electro-Mechanical Systems (MEMS) pressure transduction technology with an easy-to-use single-catheter interface. This report provides a thorough technical analysis of hardware subsystems of a device, covering topics such as implementation of embedded microcontroller-based data acquisition, design of high-precision analog signal conditioning circuits using instrumentation amplifiers, and physics of piezoresistive sensing elements. This study establishes Uropine as a feasible, scalable hardware platform that can provide tertiary-grade diagnostic precision in a portable form factor by combining theoretical biomedical engineering principles with empirical clinical data.

Downloads

Download data is not yet available.

Downloads

Published

2026-04-08

How to Cite

Singh, P., Raj, P., Tiwari, A., Pal, G. K., Thakur, V. S., Sahu, L., … Sahu, O. P. (2026). Lower Urinary Tract Symptoms Diagnosis by Uropine: A MEMS based economic urodynamic system. International Journal of Recent Advances in Engineering and Technology, 15(1), 119–126. https://doi.org/10.65521/intjournalrecadvengtech.v15i1.2065

Issue

Section

Articles

Similar Articles

<< < 17 18 19 20 21 22 23 24 25 26 > >> 

You may also start an advanced similarity search for this article.